Arbitrary Waveform Pulse Function Signal Generator FeelTech FY6900 60M DDSItem Description
frequency characteristic
Product Model
FY6900-20M
FY6900-30M
FY6900-50M
FY6900-60M
Frequency range of Sine wave
0~20MHz
0~30MHz
0~50MHz
0~60MHz
Frequency range of square wave
0~15MHz
0~20MHz
0~25MHz
0~25MHz
Frequency range of other waves
0~10MHz
0~10MHz
0~10MHz
0~10MHz
Minimum adjustable width of pulse wave
20ns
Minimum Frequency Resolution
1μHz
Waveform Length
8192 points (8K points) *14Bits
Waveform sampling rate
250MSa/s
Vertical Resolution of Waveform
14 Bits
Amplitude range (peak to peak)
frequency≤5MHz
5MHz< frequency≤10MHz
10MHz< frequency≤20MHz
frequency >20MHz
1mVpp~24Vpp
1mVpp~20Vpp
1mVpp~10Vpp
1mVpp~5Vpp
Amplitude resolution
1mV
Output impedance
50Ω±10%(typical)
Bias Adjustment Range
frequency≤20MHz:±12V; frequency >20MHz: ±2.5V
Minimum Bias Resolution
1mV
Phase adjustment range
0~359.99°
Minimum phase resolution
0.01°
display
2.42.4 inches TFT color liquid crystal display
Interface
115200 bps
Power Supply
AC100V~240V
environment condition
temperature:0~40℃ humidity:﹤80%
·Frequently Asked Questions
Fault phenomenon
Solution
Turn the power button to the "-" side, the signal generator screen is still not lit, there is no display, the channel indicator is not lit
1) Check if the power connector is connected.
2) Check that the power button is in place.
3) After completing the above checks, restart the instrument.
4) If you still cannot use the product properly, please contact
FeelElec
.
Signal generator CH2 is locked
1) Check that the signal generator is operating in sync. Press the SYSTEM button -> SYNC Enter the sync settings interface to cancel all sync parameters.
2) The power to restart the signal generator can also be unlocked.
Correct setting but no waveform output
1) Check that the BNC cable is securely connected to the corresponding [CH1] or [CH2] channel output port.
2) Check the BNC cable for internal damage.
3) Check that the BNC cable is tightly connected to the test instrument.
4) Check if the upper side indicator of the [CH1] or [CH2] button is lit. If it is not lit, press the corresponding button to make its indicator light.
5) If you still cannot use the product properly, please contact
FeelElec
.
Payment Mathod
Returns Policy
Feedback
Payment Mathod
1. Customer satisfaction is our top goal. We believe our items are so outstanding. All products are quality checked. They are new and in good condition when shipped to our customers. We are convinced you will be happy with your Purchase.
2. If product is defective or damage upon arrival, or wrong product shipped, please contact us immediately. Returns accepted within 14 days of delivery date and item must be in original new condition, not worn or altered in any way with attached tags & wrap. Otherwise deal is final. Return shipping must be paid by buyer.
3. Please contact us first if you have any problems/questions/concerns. We will be happy to resolve any issues you may have in a cordial and friendly manner.
4. We appreciate your Postive Feedback, and will do the same in return. DO NOT leave negative feedback without first communication with us. Ple
Returns Policy
1. Customer satisfaction is our top goal. We believe our items are so outstanding. All products are quality checked. They are new and in good condition when shipped to our customers. We are convinced you will be happy with your Purchase.
2. If product is defective or damage upon arrival, or wrong product shipped, please contact us immediately. Returns accepted within 14 days of delivery date and item must be in original new condition, not worn or altered in any way with attached tags & wrap. Otherwise deal is final. Return shipping must be paid by buyer.
3. Please contact us first if you have any problems/questions/concerns. We will be happy to resolve any issues you may have in a cordial and friendly manner.
4. We appreciate your Postive Feedback, and will do the same in return. DO NOT leave negative feedback without first communication
Feedback
When you satisfied with our product and services please leave us positive feedback.
If a problem occurs, contact us immediately with any email request. Just contact us using the " Ask the seller a question" link on eBay.We will maill bakc within 24hours